Services on Demand
Journal
Article
Indicators
Cited by SciELO
Related links
Similars in SciELO
Share
Revista de la Asociación Argentina de Ortopedia y Traumatología
On-line version ISSN 1852-7434
Abstract
PESCIALLO, César et al. Fractura del poste tibial en la artroplastia total de rodilla estabilizada a posterior. Presentación de dos casos. Rev. Asoc. Argent. Ortop. Traumatol. [online]. 2021, vol.86, n.1, pp.99-106. ISSN 1852-7434. http://dx.doi.org/https://doi.org/10.15417/issn.1852-7434.2021.86.1.1116.
Tibial post fracture after posterior stabilized total knee replacement (TKR) is a uncommon and disabling complication. We report two cases of non-traumatic tibial post fracture, which presented with sudden pain and recurvatum instability in joint replacements with good previous function. For the first case, the presentation time was 12 years from the primary TKR, and for the second it was 14 months. Both presented the same model of a posterior stabilized prosthesis. In both cases, a revision TKR was performed with a rotating hinge prosthesis (Endo-Model NR®, Waldemar Link), without recurrence of the deformity.
Keywords : Tibial post fracture; recurvatum; total knee replacement; posterior stabilized total knee replacement.